General
Systems is equipped for multiplexed 3D live-cell experiments:
- Multi-color, multi-position, time series, large image scan & automation
- Fully motorized high precision stage for accurate repositioning (XYZ-control)
- Stage incubator (temperature, CO2, humidity & carbogen)
- Perfusion
- Focus drift correction (Nikon PFS)
- Water immersion objectives for high-resolution on water-based samples
- Widefield epifluorescence (DAPI, CFP, GFP, YFP, mCherry & Cy5)
- Bright-field (transmitted light) for histological slides (color staining)
- Phase microscopy automated for 10x & 20x objectives
- Interference Reflection Microscopy (IRM)
- Polarized light microscopy
- FRET
- 395 nm (395/25): 40 mW
- 440 nm (438/29): 69 mW
- 470 nm (475/28): 37 mW
- 508 nm (511/16): 8 mW
- 555 nm (555/28): 108 mW
- 640 nm (635/22): 26 mW
- For fluorescent samples: sCMOS, PCO.edge camera (>80% QE, 2048×2048 pixels, 6.5 µm pixel size, 53 fps full frame, up to 500 fps in small area)
- For stained samples (histology): Nikon DS-Ri2 color CMOS camera (4908×3264 pixels, 6 fps full frame, 45 fps in 1636×1088 pixels)
- Fast LED switching (triggered acquisition)
- Fast emission filter wheel for clean multi-color imaging of DAPI, CFP, GFP, YFP, mCherry & Cy5
- Automatic water immersion dispenser (WID) for water immersion objectives
- NIS-Elements AR (Advanced Research) software including modules for: JOBS (automation), 3D Deconvolution & NIS.ai (artificial intelligence): denoise, clarify, enhance, segment & convert
Imaging Modes
LED Illumination (Lumencore, SpectraX) and max. intensities out of 40x objective
Emission Filters
Filter | λ/Bandwidth |
---|---|
Quad F67-401NXL (DAPI, GFP, mCherry, Cy5) | 435/33 520/26 595/38 695/63 |
CFP ff01-482/35-32 | 482/35 |
YFP ET540/30, F47-543 | 540/30 |
GFP ET519/26M, F47-519NXL | 519/26 |
mCherry BrightLight 595/31 | 595/31 |
Cy5 ET697/60M, F47-697NXL | 697/60 |
Detectors
Objectives
Mag. | Type | Immer. | NA | WD [μm] | Info |
---|---|---|---|---|---|
4x | PlanApo λ | Air | 0.2 | 20,000 | OFN25, ∞/- |
10x | PlanFluo Ph1 DL | Air | 0.3 | 15,200 | OFN25, ∞/1.2 |
20x | PlanApo Ph2 DM | Air | 0.75 | 1,000 | ∞/0.17 |
40x | Plan Apo λ DIC N2 | Air | 0.95 | 250-170 | OFN25, ∞/0.11-0.23 |
40x | Apo LWD λS DIC N2 | Water | 1.15 | 590-610 | ∞/0.15-0.19 |
60x | Plan Apo λ DIC N2 | Oil | 1.4 | 130 | OFN25, ∞/0.17 |
Special
Weiterführende Informationen:
https://bua.openiris.io/Landing/Res...Mögliche Nutzer:innengruppen: No restrictions
Nutzungsvoraussetzungen:
The User and PI hereby declare to have understood and accepted the AMBIO User Guidelines and Rules
including the Safety Rules, as available online at: https://ambio.charite.de/fileadmin/user_upload/microsites/ohne_AZ/sonstige/ambio/Procedures/AMBIO_User-Guidelines-Rules_2025-01-30.pdf. It is further
understood that these Guidelines and Rules may be updated or changed in agreement with the AMBIO
steering committee. Registered Users and PIs will be notified of essential changes (service fees).
Furthermore, the new User and PI have understood that a fee will be charged for the usage of AMBIO services
as outlined in the AMBIO User Guidelines and Rules, and that they are willing and able to pay the fees.
Hinweis
New users – access upon training request here.
E-Mail: ambio@charite.de
Anbieter:
Advanced Medical BIOimaging Core Facility
Standort Anbieter:
Campus Charité Mitte (CCM)
CharitéCrossOver (CCO)
Rm 01-306
Virchowweg 6
10117 Berlin
Germany
Standort:
Charité Campus Mitte
CCO Rm. 01-307
Virchowweg 6
10117 Berlin
Germany